Lessons from the Lab: Cyanide Testing Requirements

In our blog posts Lessons From the Lab we answer frequently asked questions from clients.  Find all installments of Lessons From the Lab here

Cyanide sampling requirements have become stricter over the years. The Nevada Division of Environmental Protection (NDEP) issued guidance in October of 2015 that cyanide analysis must be collected correctly in the field so as not to have samples rejected by the analytical laboratory, or by the state due to incorrect sampling procedures.

NDEP stated, “If you are analyzing Cyanide samples for compliance with a Nevada program, (SDWA, CWA, RCRA, Mining) samples must be collected as described below (ASTM D-7365-09).  Data obtained from samples not collected as described in ASTM D-7365-09 will be rejected.”

“ASTM D-7365-09 8.2.1 states that sample containers shall be made of materials that will not contaminate the sample, cleaned thoroughly to remove all extraneous surface contamination prior to use.  Chemically resistant glass containers as well as rigid plastic containers made of high density polyethylene (HDPE) are suitable.  Samples should be collected and stored in amber gas tight vials or narrow mouth bottles to minimize exposure to ultraviolet radiation and to minimize headspace in the sample containers (for example, amber open top VOA vials, amber Boston round bottles, or amber narrow-mouth HDPE bottles).”

“All certified Laboratories must reject samples not collected in suitable containers.”

What does this mean? All samples, regardless of matrix (drinking water, wastewater, ground water, surface water, aqueous, soil, sludge, etc.), must be collected in an amber narrow mouth container to minimize UV radiation exposure and to minimize headspace in sample containers.  Samples not collected in the correct containers must be rejected by the laboratory and the sample should be collected in the correct containers, as described above. Furthermore, as dictated by the method cited by NDEP, chemical preservation is also required for aqueous samples.  Aqueous samples must be preserved with sodium hydroxide (NaOH) to pH >10 at the time of collection, and then chilled on ice.

At WETLAB, we provide the appropriate bottles and preservative (NaOH) needed for your cyanide analysis, and are happy to answer any questions you may have regarding cyanide sampling containers.

Please call us at any at 775-355-0202 to request sample containers.